14
5. When the desired value is selected,
press
ENTER
to save and progress to
the next
MENU
level.
At any
Menu
level, pressing
ENT
twice will
return to the Main Display Screen at the
previously set
Operating Mode.
(2-4) End Timer
The Auditor™ family of products shares
software with other items in the Auditor™
line. The End Timer has no purpose with
the
AUET
products.
Settings in this screen have no effect
on the operation of the tester.
Press
ENTER
to progress to the next
MENU
level.
At any
Menu
level, pressing
ENT
twice will
return to the Main Display Screen at the
previously set
Operating Mode
(2-5) Sign Lock
The
SIGN LOCK
feature allows the user
to select the rotation direction in which a
torque peak will be captured. With
SIGN
LOCK
set to
ON
, the initial first rundown
direction is set as the default direction.
If the initial rundown was performed in a
clockwise direction a positive reading
indicates a reading taken in the clockwise
direction while a negative torque value
indicates a reading taken in the counter-
clockwise direction. If the initial rundown is
taken in a counter clockwise manner, then
the opposite will hold true. Any readings
taken in the non-selected direction will be
measured, but not captured / displayed as
a peak value.
With
SIGN LOCK
turned
OFF
, peak
values in both the clockwise and counter-
clockwise directions will be captured and
displayed.
Since
most
torque
rundowns
are
conducted and readings are taken in the
clockwise direction, the user may find it
beneficial to turn this feature
ON
.
Turning Sign Lock ON / OFF
1. From the main display press
MENU
2. Press the
Right Selection
button to
select
PEAK OPTIONS MENU
3. Use the
Selection Arrow
buttons to
select
SIGN LOCK
4. Use the
Left Selection button
to
toggle
SIGN LOCK
on or off.
5. When the desired value is selected,
press
ENTER
to save and progress to
the next
MENU
level.
At any
Menu
level, pressing
ENT
twice will
return to the Main Display Screen at the
previously set
Operating Mode
Resetting Sign Lock
If
SIGN LOCK
is
ON
and the rotation
direction needs to be reset turn the
AUET
OFF
then turn the
AUET
back on and
follow the above procedure.
(2-6) Edit Limits
By using the
EDIT LIMITS
feature to set
high and low limits for torque readings, the
user
can
receive
visual
feedback
regarding the rundown results.
Setting the
HIGH
and
LOW
limits
appropriately will result in the following
display signals:
Reading
outside
of set limits:
A
RED
LED being lit on the
AUET
(
Next to
OFF
button)
A
symbol for
LOW
readings and a
symbol for
HIGH
readings being
displayed on the screen
(Figure 1)
Reading
within
set limits:
A
GREEN
LED being lit on the
AUET
(
Next to
ON
button
)
(Figure 1
)
Setting High and Low Torque Limits
1. From the main display press
MENU.
2. Press the
Right Selection
button to
select
PEAK OPTIONS MENU.
3. Use the
Selection Arrow
buttons to
select
EDIT LIMITS.