QuickSwap Installation and Startup
Overview
Final leak test procedure for EI sources
This procedure applies to MSDs with an EI source. For MSDs with a CI source
installed, see
“Final leak test procedure for CI sources”
1
From the
Instrument Control
screen, select
Instrument
, then
Edit Tune
Parameters
. This brings up the
Manual Tune
2
Select the
PFTBA
field. Move the
PFTBA
slider at the lower right to the
CLOSED
position. Confirm that the
PFTBA
field now reads
CLOSE
.
Figure 13
Manual Tune scan in progress. Air is high due to recent venting and
pumpdown