QuickSwap Installation and Startup
Update Auxiliary EPC Controller Firmware
11
In the bottom right corner, configure and set the EPC module firmware. See
a
Select the appropriate
Aux channel
(default: AUX 5).
b
Select
QuickSwap
from the PID Settings drop down list.
c
Click
Set
to download the new PID constants into the auxiliary EPC
module.
12
Restart the instrument when prompted.
Firmware update is now complete, and the auxiliary EPC module is ready for
use with QuickSwap.
Should you ever wish to use the auxiliary channel for an application other
than QuickSwap, you can revert the firmware to its factory state using the
firmware update utility. Follow the procedure in this section but in
,
select
Default
from the PID Settings drop down list.
Figure 5
Firmware update utility connected to an instrument