Overview
QuickSwap Installation and Startup
19
5
For users who plan to use a CI source (CI MSDs only), check the MSD
ChemStation version.
If using MSD ChemStation version
D.01.xx or earlier
, install the EI source.
Agilent recommends using EI mode for easier installation and leak testing
in this case.
If using MSD ChemStation version
D.02.xx or later
, proceed with either the
EI or CI source installed.
“Final leak test procedure for CI sources”
2. Select and Install a Restrictor
1
Install the lock nut on the end of the MSD transfer line (
) with the
threads facing the MSD. Run it down until it touches the transfer line (stops
turning). Do not tighten it.
2
Restrictors are precut lengths of deactivated fused-silica tubing with
swaged ferrules attached.
•
For initial installation, Agilent recommends installing a 100-µm
restrictor for use with the checkout methods supplied on the QuickSwap
User Information CD.
•
If not using one of these methods, or if this is not an initial installation,
refer to the
Reference Manual
to choose an appropriate restrictor.
Figure 6
Installing the lock nut
Transfer line
Lock nut
Threads engaged
Touching transfer line