![Agilent Technologies QuickSwap G3185B Installation And Setup Download Page 26](http://html1.mh-extra.com/html/agilent-technologies/quickswap-g3185b/quickswap-g3185b_installation-and-setup_2867789026.webp)
26
QuickSwap Installation and Startup
Overview
3
Find the appropriate SilTite metal ferrule and internal nut from
.
You will also need two 1/4-inch open end wrenches and a column cutter.
4
Thread the column end through the internal nut and the ferrule (tapered
end away from the nut) leaving approximately 3 cm of column protruding
beyond the ferrule. Thread the swaging nut onto the internal nut with the
column protruding (
).
Table 2
Available metal ferrule and nut packages
Part number
Ferrule description
Quantity
Column id, µm
5188-5361
0.4-mm id
pkg of 10
100, 250
5188-5362
0.5-mm id
pkg of 10
320
5188-5383
0.8-mm id
pkg of 10
530
C A U T I O N
Wear clean, lint-free gloves when handling the column end and fittings to avoid
contamination.
Figure 11
Thread the internal nut and ferrule onto column and assemble with swaging
nut
Internal nut
Swaging nut
Column
Ferrule