92
Agilent 1290 Infinity Binary Pump User Manual
7
Error Information
Pump Error Messages
Overcurrent of solvent selection valve (SSV)
Error ID: 22412
Overcurrent of solvent selection valve (SSV).
•
Parameter: None
Deliver underrun
Error ID: 22413
Internal error.
•
Parameter: None
Defect connection between main board and pump drive encoder
Error ID: 22414
Defect connection between main board and pump drive encoder.
•
Parameter: 1 – 4 referring to pump drive
Probable cause
Suggested actions
1
Valve defect.
Replace the solvent selection valve.
Probable cause
Suggested actions
1
Internal error.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Defect connection between main board and
pump drive encoder.
Please contact your Agilent service
representative.
2
Defect encoder.
Please contact your Agilent service
representative.