Agilent 1290 Infinity Binary Pump User Manual
39
Installing the Module
3
Installation of seal wash option
Installation of seal wash option
Figure 9
Binary Pump with Seal Wash Function
The 1290 Infinity Binary Pump is optionally available with a seal wash
function. This option is recommended when using buffers or other non-volatile
solvents or additives that could deposit on pistons and seals. It is used for
regularly cleaning these parts automatically.
1
Place a wash solvent reservoir into the solvent cabinet. A mixture of
distilled water and isopropanol (90/10) is a good choice for many
applications.
2
Put the solvent inlet tube into the solvent reservoir, close it and connect the
tube to the seal wash pump.
3
Route the outlet of the wash tube into a waste container.