Agilent 1290 Infinity Binary Pump User Manual
115
Maintenance
9
Installing Fittings and Capillaries
Installing Fittings and Capillaries
1
Install fittings and capillaries.
2
Tighten fittings and capillaries.
WA R N I N G
Solvent can spray under high pressure.
➔
Observe appropriate safety procedures (for example, goggles, safety gloves and
protective clothing), when opening flow path.
C A U T I O N
Deformation of fittings and seals
Liquid drops under high pressure up to 1200 bar act like solid parts. Tightening
connections under high pressure can deform or destroy fittings and seals.
➔
Never tighten flow connections under pressure.
N O T E
The lifetime of a fitting depends on how firmly it has been tightened; firm tightening
reduces the lifetime.
If fitting has been overtightened, replace it.