LAN Configuration
Initialization mode selection
Initialization mode selection
The following initialization (init) modes are selectable:
Bootp
When the initialization mode
Bootp
is selected, the module tries to download
the parameters from a
Bootp
Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the module. Therefore, the parameters are lost with the next power
cycle of the module.
Figure 37
Bootp
(Principle)
Table 16
Initialization Mode Switches
SW 6
SW 7
SW 8
Init Mode
OFF
OFF
OFF
Bootp
OFF
OFF
ON
Bootp & Store
OFF
ON
OFF
Using Stored
OFF
ON
ON
Using Default
ON
OFF
OFF
DHCP
1
1
Requires firmware B.06.40 or above. Modules without LAN on board, see G1369C LAN Interface Card