Agilent 1290 Infinity Binary Pump User Manual
9
Maintenance
Releasing a Stuck Inlet Valve
3
Attach a Luer lock syringe with adapter to the tubing and
fill it with solvent.
4
Reconnect tubing to inlet valve.
5
Unscrew tubing at degassing unit and attach the syringe
to it.
6
Push solvent with syringe until it comes out at the top of
the High Pressure Filter Assembly.