Site Requirements and Specifications
Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Consideration
The module power supply has wide ranging capabilities and accepts any line
voltage in the range mentioned in
on page 19. Consequently, there is
no voltage selector in the rear of the module. There are also no externally
accessible fuses, because automatic electronic fuses are implemented in the
power supply.
WA R N I N G
Module is partially energized when switched off, as long as the power cord is
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
➔
Make sure that it is always possible to access the power plug.
➔
Remove the power cable from the instrument before opening the cover.
➔
Do not connect the power cable to the Instrument while the covers are removed.
WA R N I N G
Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are connected
to line voltage higher than specified.
➔
Connect your module to the specified line voltage.