Hardware Information
12
Firmware Description
Firmware Updates
Firmware updates can be done using your user interface:
•
PC and Firmware Update Tool with local files on the hard disk
•
Instant Pilot (G4208A) with files from a USB Flash Disk
•
Agilent Lab Advisor software B.01.03 and above
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315AB for the G1315A/B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
VVV is the revision number, for example 102 is revision 1.02,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section
Replacing Firmware
in
chapter
"Maintenance"
or use the documentation provided with the
Firmware Update Tools
.
Figure 30
Firmware Update Mechanism
N O T E
Update of main system can be done in the resident system only. Update of the resident
system can be done in the main system only.
Main and resident firmware must be from the same set.