Agilent 1290 Infinity Binary Pump User Manual
237
LAN Configuration
13
Automatic configuration with Bootp
Automatic configuration with Bootp
About Agilent BootP Service
The Agilent BootP Service is used to assign the LAN Interface with an IP
address.
The Agilent BootP Service is provided on the ChemStation DVD. The Agilent
BootP Service is installed on a server or PC on the LAN to provide central
administration of IP addresses for Agilent instruments on a LAN. The BootP
service must be running TCP/IP network protocol and cannot run a DHCP
server.
N O T E
All examples shown in this chapter will not work in your environment. You need your own
IP-, Subnet-Mask- and Gateway addresses.
N O T E
Assure that the detector configuration switch is set properly. The setting should be either
BootP
or
BootP & Store
, see
N O T E
Assure that the detector connected to the network is powered off.
N O T E
If the Agilent BootP Service program is not already installed on your PC, then install it from
your Agilent ChemStation DVD, located in folder
BootP
.