30-49
Waveform Commands
PREamble
Table 30-1
Waveform Preamble Elements
Element
Description
Format
The format value describes the data transmission mode for waveform data
output. This command controls how the data is formatted when it is sent
from the oscilloscope. (See :WAVeform:FORMat.)
Type
This value describes how the waveform was acquired.
(See also the :WAVeform:TYPE? query.)
Points
The number of data points or data pairs contained in the waveform data.
(See :ACQuire:POINts.)
Count or Segments For the AVERAGE waveform type, the count value is the minimum count or
fewest number of hits for all time buckets. This value may be less than or
equal to the value requested with the :ACQuire:AVERage:COUNt command.
For NORMAL, RAW, INTERPOLATE, and VERSUS waveform types, this value
is 0 or 1. The count value is ignored when it is sent to the oscilloscope in the
preamble. (See :WAVeform:TYPE and :ACQuire:COUNt.)
Segments is used instead of Count when the data is acquired using the
Segmented memory acquisition mode. This number is the total number of
segments that were acquired.
X Increment
The X increment is the duration between data points on the X axis.
For time domain waveforms, this is the time between points.
(See the :WAVeform:XINCrement? query.)
X Origin
The X origin is the X-axis value of the first data point in the data record.
For time domain waveforms, it is the time of the first point. This value is
treated as a double precision 64-bit floating point number.
(See the :WAVeform:XORigin? query.)
X Reference
The X reference is the data point associated with the X origin. It is at this
data point that the X origin is defined. In this oscilloscope, the value is
always zero. (See the :WAVeform:XREFerence? query.)
Y Increment
The Y increment is the duration between Y-axis levels. For voltage
waveforms, it is the voltage corresponding to one level.
(See the :WAVeform:YINCrement? query.)
Y Origin
The Y origin is the Y-axis value at level zero. For voltage waveforms, it is the
voltage at level zero. (See the :WAVeform:YORigin? query.)
Y Reference
The Y reference is the level associated with the Y origin. It is at this level
that the Y origin is defined. In this oscilloscope, this value is always zero.
(See the :WAVeform:YREFerence? query.)
Coupling
The input coupling of the waveform. The coupling value is ignored when
sent to the oscilloscope in the preamble. (See the :WAVeform:COUPling?
query.)
X Display Range
The X display range is the X-axis duration of the waveform that is displayed.
For time domain waveforms, it is the duration of time across the display.
(See the :WAVeform:XRANge? query.)
Summary of Contents for Infiniium 8000A
Page 1: ...Agilent Technologies Infiniium 8000A Programmer s Reference ...
Page 2: ......
Page 20: ...Contents 16 ...
Page 21: ...1 Introduction to Programming ...
Page 43: ...2 LAN and GPIB Interfaces ...
Page 53: ...3 Message Communication and System Functions ...
Page 58: ...3 6 ...
Page 59: ...4 Status Reporting ...
Page 78: ...4 20 Figure 4 3 Status Reporting Decision Chart ...
Page 79: ...5 Programming Conventions ...
Page 84: ...5 6 Programming Conventions The Command Tree Figure 5 1 Command Tree ...
Page 85: ...5 7 Programming Conventions The Command Tree Figure 5 2 Command Tree Continued ...
Page 86: ...5 8 Programming Conventions The Command Tree Figure 5 3 Command Tree Continued ...
Page 87: ...5 9 Programming Conventions The Command Tree Figure 5 4 Command Tree Continued ...
Page 88: ...5 10 Programming Conventions The Command Tree Figure 5 5 Command Tree Continued ...
Page 89: ...5 11 Programming Conventions The Command Tree Figure 5 6 Command Tree Continued ...
Page 94: ...5 16 ...
Page 95: ...6 Sample Programs ...
Page 149: ...7 Acquire Commands ...
Page 176: ...7 28 Acquire Commands SRATe AUTO ...
Page 177: ...8 Bus Commands ...
Page 187: ...9 Calibration Commands ...
Page 195: ...10 Channel Commands ...
Page 223: ...11 Common Commands ...
Page 247: ...12 Digital Commands ...
Page 254: ...12 8 ...
Page 255: ...13 Disk Commands ...
Page 300: ...13 46 Disk Commands STORe Obsolete ...
Page 301: ...14 Display Commands ...
Page 322: ...14 22 ...
Page 323: ...15 External Trigger Commands ...
Page 343: ...16 Function Commands ...
Page 382: ...16 40 ...
Page 383: ...17 Hardcopy Commands ...
Page 391: ...18 Histogram Commands ...
Page 403: ...19 InfiniiScan ISCan Commands ...
Page 421: ...20 Limit Test Commands ...
Page 429: ...21 Marker Commands ...
Page 452: ...21 24 ...
Page 453: ...22 Mask Test Commands ...
Page 499: ...23 Measure Commands ...
Page 636: ...23 138 Measure Commands VUPPer ...
Page 637: ...24 Pod Commands ...
Page 642: ...24 6 ...
Page 643: ...25 Root Level Commands ...
Page 645: ...25 3 STORe SETup STORe WAVeform TER Trigger Event Register VIEW ...
Page 674: ...25 32 ...
Page 675: ...26 Self Test Commands ...
Page 679: ...27 System Commands ...
Page 694: ...27 16 ...
Page 695: ...28 Time Base Commands ...
Page 708: ...28 14 ...
Page 709: ...29 Trigger Commands ...
Page 822: ...29 114 ...
Page 823: ...30 Waveform Commands ...
Page 893: ...31 Waveform Memory Commands ...
Page 902: ...31 10 ...
Page 903: ...32 Error Messages ...
Page 914: ...32 12 ...