Setting the Clock
2-2
Preparing Your H
EARTSTREAM
FR2 for Use
The display screen brings up the main menu. From this menu, you can start the FR2
battery insertion selftest, review information from the last time the FR2 was used, or
go to the next screen for other options. Information about the optional data card and
the battery status is also provided. (See Chapter 7, Data Management and Review,
for details about reviewing an incident and using a data card.)
NOTE:
This screen will
not
be displayed if the FR2 is connected to defibrillation
pads (that are applied to the patient) when the battery is inserted, and you will not
be able to access the menu items. In addition, the battery insertion selftest and
periodic automatic selftests cannot run while the defibrillation pads are connected.
Be sure to unplug the pads connector from the FR2 after each use. Do not store
the FR2 with the pads connected.
NOTE
:
To move around the menus displayed, use the Option buttons as follows:
•
Press the LOWER Option button to move the highlight bar from one item to
another on the menu.
•
Press the UPPER Option button to select the highlighted item or to scroll
through the settings for that item.
If you select NEXT, the menu displayed lets you review the history of the FR2, review
the history of the battery being used, access setup data, set the clock, or return to
the first menu. (See Chapter 4, Maintaining, Testing, and Troubleshooting Your
H
EARTSTREAM
FR2, for details about the review options and Chapter 6, Setup and
Advanced Mode Features, for information on the setup option.)
NOTE:
If you make no selection for 10 seconds, the selftest will automatically
run.
If you want to select something different from either of these menus, you must
do so before the selftest begins, or remove and reinstall the battery to bring up the
main menu. You can press the On/Off button at any time to turn off the FR2 and
return it to standby (ready for use) mode. To use the FR2 when it is in standby mode,
press the On/Off button to turn it on.
Setting the Clock
It is recommended that the first time you prepare your H
EARTSTREAM
FR2 for use,
you check the FR2’s internal clock to be sure it is set to the correct date and local
time. You can reset it if necessary.
To see the clock settings, select NEXT from the first menu, within 10 seconds of
installing the battery, and then select CLOCK. To do this:
RUN SELFTEST
REVIEW INCIDENT
NO DATA CARD
GOOD BATTERY
NEXT
IN EMERGENCY
PRESS OFF TO QUIT
DEVICE HISTORY
BATTERY HISTORY
SETUP
CLOCK
RETURN
IN EMERGENCY
PRESS OFF TO QUIT
RUN SELFTEST
REVIEW INCIDENT
NO DATA CARD
GOOD BATTERY
NEXT
IN EMERGENCY
PRESS OFF TO QUIT
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...