Troubleshooting Guide
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
4-13
4
Flashing
red
X
Audio signal:
chirping
• Battery power is low.
• The FR2 has been stored outside
the recommended temperature
range.
• An error has been detected as
part of the selftest.
• The FR2 has been unable to
perform its daily selftests.
• Remove and reinstall the battery and
run a battery insertion selftest. A
screen prompt will tell you if the FR2
has been stored outside the
recommended temperature range. See
Appendix B for recommended range.
• Remove and reinstall the battery and
perform the battery insertion selftest.
If it fails, install a new battery and
repeat the test. If it fails again, contact
Agilent Technologies for technical
support.
• Make sure defibrillation pads are not
attached to the FR2.
Solid red
X
None
• The Training & Administration
Pack is being used in the
administration function; solid red
X
is normal.
• The battery is missing or
completely depleted.
• Remove the Training &
Administration Pack and install the
battery.
• Remove and reinstall the battery and
perform the battery insertion selftest.
If it fails, install a
new
battery and
repeat the test. If it fails again, contact
Agilent Technologies for technical
support.
STATUS
INDICATOR
OTHER
SIGNAL
POSSIBLE CAUSE
RECOMMENDED ACTION
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...