Testing
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
4-7
4
Screen prompts guide you through a series of steps in the interactive part of the
selftest. Some ask you to observe that a feature of the H
EARTSTREAM
FR2 works
properly. Others ask you to take certain actions — for example, to press a button. The
screen then displays a message showing that the button’s operation has been
verified. If you do not press the button, or if you do but the button is not working,
the screen displays a message that the button’s function is not verified.
It is important for you to verify correct operation of each item tested. If something
does not work correctly — for example, if lights do not come on or you do not hear
beeps when expected — make a note of the problem and contact Agilent
Technologies.
NOTE:
Do not use the H
EARTSTREAM
FR2 until
all
parts of the interactive selftest
verify correct performance. Be sure to note and report any problems you find.
The following table describes the parts of the H
EARTSTREAM
FR2 tested in the
interactive part of the selftest and any action you are asked to take.
SHOCK BUTTON
VERIFIED
IN EMERGENCY
PRESS OFF TO QUIT
PRESS THE
SHOCK BUTTON
IN EMERGENCY
PRESS OFF TO QUIT
CHECK
SHOCK BUTTON LIGHT
AND
PADS CONNECTOR LIGHT
IN EMERGENCY
PRESS OFF TO QUIT
FEATURE
TEST DESCRIPTION
Speaker/Beeper
Screen prompt: CHECK SPEAKER SOUND (2 beeps)
Listen for the two beeps, one from the beeper and then one from the speaker.
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...