Testing
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
4-9
4
If proper operation of all features has not been verified in the interactive selftest,
you may want to rerun the battery insertion selftest. If a feature of operation cannot
be verified, contact Agilent for technical support.
Periodic Selftests
In addition to the battery insertion selftest, the FR2 automatically performs periodic
selftests (PSTs). These daily, weekly, and extensive monthly selftests check many
important functions of the FR2, including battery capacity and internal circuitry.
If it detects a problem during one of these periodic selftests, the FR2 beeps and
displays a flashing red
X
or a solid red
X
on the Status Indicator.
Device History
The H
EARTSTREAM
FR2 stores key information about its history in internal memory.
To review the history of your FR2, select NEXT from the menu screen displayed
when you insert the battery, then select DEVICE HISTORY from the next menu
displayed.
The device history information is automatically stored in the internal memory of the
H
EARTSTREAM
FR2. It includes:
USES — how many times the H
EARTSTREAM
FR2 has been used (shown in the left
column of numbers) and the total time in minutes it has been used (shown in
the right column of numbers);
SHOCKS — the total number of shocks it has delivered;
TRAINING — how many times it has been used with the Training &
Administration Pack for training (left column) and the total time in minutes it
has been used for training (right column); and
TESTS — how many tests have been run. Four figures are shown: daily (upper
left), weekly (upper right), and monthly (lower left) periodic selftests, and
battery insertion selftests (lower right).
Battery History
Information about use of the battery currently installed in your H
EARTSTREAM
FR2 is
also available. To review the history of the battery, select NEXT from the menu
screen displayed when you insert the battery, then select BATTERY HISTORY from
the next menu displayed.
The battery history information is automatically stored in the internal memory of
the battery. It includes:
USE MINUTES — the total operating time (in minutes), including selftest time,
for this battery;
DEVICE HISTORY
BATTERY HISTORY
SETUP
CLOCK
RETURN
IN EMERGENCY
PRESS OFF TO QUIT
DEVICE HISTORY
RETURN
USES:
12
19
SHOCKS:
17
TRAINING:
25
456
TESTS:
156
22
5
1095
DEVICE HISTORY
BATTERY HISTORY
SETUP
CLOCK
RETURN
IN EMERGENCY
PRESS OFF TO QUIT
BATTERY HISTORY
RETURN
USE MINUTES:
519
CHARGES:
40
GOOD BATTERY
STATUS:
00000000
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...