F-2
Clinical Summary
The 150 J SMART Biphasic waveform defibrillated 98% of VF patients in the first
series of three shocks or fewer compared with 69% of patients treated with
monophasic waveform shocks. Outcomes are summarized as follows:
Conclusions
The 150 J SMART Biphasic waveform defibrillated at higher rates than the 200-360 J
monophasic waveforms, resulting in more patients achieving return of spontaneous
circulation (ROSC) (p=0.01). EMS system outcomes of survival discharge were not
significantly different statistically. However, patients resuscitated with the
lower-energy SMART Biphasic waveform were more likely to have good cerebral
performance (CPC, cerebral performance category) (p=0.04).
SMART Biphasic Patients
Number (%)
Monophasic Patients
Number (%)
P Value
(chi square)
Defibrillation Efficacy
Single shock only
</= 2 shocks
</= 3 shocks
52/54 (96%)
52/54 (96%)
53/54 (98%)
36/61 (59%)
39/61 (64%)
42/61 (69%)
<0.0001
<0.0001
<0.001
Patients Defibrillated
54/54 (100%)
49/58 (84%)
0.003
ROSC
41/54 (76%)
33/61 (54%)
0.01
Survival to Hospital Admission
33/54 (61%)
31/61 (51%)
0.27
Survival to Hospital Discharge
15/54 (28%)
19/61 (31%)
0.69
CPC = 1 (Good)
13/15 (87%)
10/19 (53%)
0.04
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...