M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
B-3
B
H
EARTSTREAM
FR2 AED Specifications
ECG Analysis System
“Charge Complete”
Indicator
Shock button flashes, audio tone sounds.
Disarm (AED mode)
Once charged, the H
EARTSTREAM
FR2 will dump the charge if:
• if patient’s heart rhythm changes to non-shockable rhythm,
OR
• a shock is not delivered within 30 s after the FR2 is armed,
OR
• the PAUSE button (if enabled) is pressed,
OR
• the On/Off button is pressed to turn off the FR2,
OR
• the defibrillation pads are removed from the patient or the pads connector is
disconnected from the FR2.
Disarm (advanced mode)
Once charged, the H
EARTSTREAM
FR2 will dump the charge if:
in advanced mode ANALYZE
• the MANUAL DISARM button is pressed,
OR
• if patient’s heart rhythm changes to non-shockable rhythm,
OR
• a shock is not delivered within 30 s after the FR2 is armed,
OR
• the On/Off button is pressed to turn off the FR2,
OR
• the defibrillation pads are removed from the patient or the pads connector
is disconnected from the FR2.
in advanced mode CHARGE (M3860A only)
• the MANUAL DISARM button is pressed,
OR
• a shock is not delivered within 30 s after charging ,
OR
• the On/Off button is pressed to turn off the FR2,
OR
• the defibrillation pads are removed from the patient or the pads connector
is disconnected from the FR2.
Shock Delivery Vector
Via defibrillation pads placed in the anterior-anterior (Lead II) position.
CATEGORY
NOMINAL SPECIFICATIONS
CATEGORY
NOMINAL SPECIFICATIONS
Function
Evaluates impedance of defibrillation pads for proper contact with patient skin,
and evaluates the ECG rhythm and signal quality to determine if a shock is
appropriate.
Protocols
Follows pre-programmed settings to match local EMS guidelines or medical
protocols. The settings can be modified using the setup options.
Summary of Contents for Heartstream FR2
Page 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Page 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Page 9: ...iv Contents Notes...
Page 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Page 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Page 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Page 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Page 65: ...Accessories A 2 Notes...
Page 77: ...Symbols and Controls D 4 Notes...