Installation Note E8364-90023
Step 7. Remove the Old Bottom Side Hardware and Cables
•
Remove the Existing Configurable Test Set Cables
•
•
•
Remove the A22 Switch/Splitter
•
Remove the A23 and A24 Detectors
•
Remove the A27, A28, A29, and A30 First Converters (Mixers)
•
Remove the A25 and A26 Test Port Couplers
•
Remove the A20 LODA (5087-7129)
Remove the Existing Configurable Test Set Cables
1. Using a 5/16-inch wrench, remove the hex nuts from all 12 front panel connectors (item
①
). Remove the lock washers. Retain both the hex nuts and lock washers for installation of
new connectors later.
2. Using a 5/16-inch wrench, remove each of the configurable test set cables (item
②
) from
the analyzer. Discard these cables, they will be replaced with new ones.
Figure 9
Existing Configurable Test Set Cables Removal