Installation Note E8364-90023
15
NOTE
The previous figure and the figure below show the E8364B front panel and
floppy disk drive. The E8363C and E8364C front panel have a slightly
different appearance and does not include a floppy disk drive.
Remove the A12 Source 20
1. With a 5/16-inch wrench, disconnect semirigid cable (item
①
).
2. Disengage the A12 source 20 from the A11 connector and slide it out of the analyzer while
observing the
CAUTION
below.
CAUTION
To avoid damaging the semirigid cable (item
①
), pull down on the cable from the
analyzer’s bottom side to move it away while sliding out the A12 source 20. The
cable has some spring to it – the RF pin could be damaged by the A12 source 20.
Figure 5
A12 Source 20 Removal