44
Installation Note E8364-90023
Step 13. Enter the New Model Number
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A keyboard and a mouse must be connected to the analyzer for this procedure.
Enter the New Model Number
1. On the analyzer’s
System
menu, point to
Service
, then click
Option
Enable
.
2. Click
Repair
.
3. Check
Model shown is incorrect!
, then click
Begin Repair
.
4. In the
Model Change
box, select the new model number,
E8361A
or
E8361C
(the E8363B
and E8364B upgrade to an E8361A; the E8363C and E8364C upgrade to an E8361C).
5. Using the keyboard, enter the keyword obtained in
“Step 1. Obtain a Keyword and Verify
in the box provided. Enter this keyword
exactly
as provided.
6. Click
Enter Model
.
7. Verify that the new model number displayed is correct, then click YES. BE CAREFUL.
Once you click YES the entry cannot be changed without first obtaining another keyword
from Agilent.
If it is not correct, click
NO
and repeat the repair.
8. When the installation is complete, click
Exit
.
Verify that the New Model Number is Shown Correctly
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that the new model number, E8361A or E8361C, is shown correctly in the display.
3. Click OK.
4. On the analyzer front panel, press
Preset
. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 67 GHz.