![Agilent Technologies E8257D User Manual Download Page 281](http://html.mh-extra.com/html/agilent-technologies/e8257d/e8257d_user-manual_2867504281.webp)
Chapter 12
267
Peripheral Devices
N5102A Digital Signal Interface Module
Configuring the Clock Signal
1.
Press the
Clock Setup
softkey, as shown in
Figure 12- 17
.
Figure 12-17
Clock Setup Menu Location
From this softkey menu, set all of the clock parameters that synchronize the data between the
N5102A module and the device. From this menu, the clock signal phase can be changed so the
clock occurs during the valid portion of the data.
Figure 12- 18
shows the clock setup menu.
If the device or external clock does not match the frequency, one of the following error messages
will appear on the PSG:
803
Digital module input FIFO overflow error; There are more
samples being produced than can be consumed at the current
clock rate. Verify that the digital module clock is set up
properly.
This error is reported when the digital module clock setup is not
synchronized with the rate the samples are entering the digital
module. Verify that the input clock rate matches the specified
clock rate under the clock setup menu.
804
Digital module input FIFO underflow error; There are not enough
samples being produced for the current clock rate. Verify that
the digital module clock is set up properly.
This error is reported when the digital module clock setup is not
synchronized with the rate the samples are entering the digital
module. Verify that the input clock rate matches the specified
clock rate under the clock setup menu.
Accesses the Clock Setup Menu
Summary of Contents for E8257D
Page 12: ...Contents xii ...
Page 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Page 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Page 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Page 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...