![Agilent Technologies E8257D User Manual Download Page 107](http://html.mh-extra.com/html/agilent-technologies/e8257d/e8257d_user-manual_2867504107.webp)
Chapter 3
93
Basic Digital Operation
Using the Dual ARB Waveform Player
2.
Create the first waveform segment:
a.
Press
Mode
>
Dual ARB
>
Waveform Segments
>
Load Store
to Store.
b.
Highlight the default segment
AUTOGEN_WAVEFORM
.
c.
Press
Rename Segment
>
Editing Keys
>
Clear Text
.
d. Enter a file name (for example,
TTONE
), and press
Enter
>
Store Segment To NVWFM Memory
.
This renames the waveform segment, and stores a copy in non volatile memory.
3.
Generate the second waveform:
a.
Press
Mode
>
Multitone
>
Initialize Table
>
Number Of Tones > 9 > Enter > Done
.
b.
Press
Multitone Off On
to On, then Off.
Remember that a waveform cannot be renamed as a segment while it is in use.
This generates a multitone waveform with nine tones. During waveform generation, the
M–TONE
and
I/Q
annunciators activate. The waveform is stored in volatile memory with the default file
name
AUTOGEN_WAVEFORM
.
4.
Create the second waveform segment:
Repeat Step 2, giving this segment a descriptive name (for example,
MTONE
).
Building and Storing a Waveform Sequence
This example shows how to build and edit a waveform sequence using the two waveform segments
created on
page 92
. To use a segment in a sequence, the segment must reside in volatile memory; for
information on loading waveform segments from non volatile to volatile memory, see
page 95
.
1.
Select the waveform segments:
Define a sequence as one repetition of the two–tone waveform segment followed by one repetition
of the nine–tone multitone waveform segment.
a.
Press
Mode
>
Dual ARB
>
Waveform Sequences
>
Build New Waveform Sequence
>
Insert Waveform
.
b.
Highlight the a waveform segment (for example,
TTONE
) and press
Insert Selected Waveform
.
c.
Highlight a second waveform segment (for example,
MTONE
) and press
Insert Selected Waveform
.
d. Press
Done Inserting
2.
Optional
: Enable markers as desired for the segments in the new sequence: see
page 106
.
3.
Name and store the waveform sequence to the
Catalog of Seq Files
in the memory catalog:
a.
Press
Name and Store
.
b.
Enter a file name (for example,
TTONE+MTONE
).
c.
Press
Enter
.
Summary of Contents for E8257D
Page 12: ...Contents xii ...
Page 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Page 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Page 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Page 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...