![Agilent Technologies E8257D User Manual Download Page 140](http://html.mh-extra.com/html/agilent-technologies/e8257d/e8257d_user-manual_2867504140.webp)
126
Chapter 3
Basic Digital Operation
Using Waveform Scaling
Figure 3-18
Waveform Overshoot
How Scaling Eliminates DAC Over–Range Errors
Scaling reduces or shrinks a baseband waveform’s amplitude while maintaining its basic shape and
characteristics, such as peak–to–average power ratio. If the fast–rising baseband waveform is scaled
enough to allow an adequate margin for the overshoot, the interpolator filter is then able to calculate
sample points that include the ripple effect, thereby eliminating the over–range error
(see
Figure 3- 19
).
Figure 3-19
Waveform Scaling
Summary of Contents for E8257D
Page 12: ...Contents xii ...
Page 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Page 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Page 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Page 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...