238
Chapter 11
AWGN Waveform Generator
Configuring the AWGN Generator
Generating the Waveform
Press
AWGN Off On
until On is highlighted.
This generates an AWGN waveform with the parameters defined in the previous procedure. During
waveform generation, the
AWGN
and
I/Q
annunciators activate and the AWGN waveform is stored in
volatile ARB memory. The waveform is now modulating the RF carrier.
Real Time I/Q Baseband AWGN
1.
Press
Preset
.
2.
Press
Mode
>
More (1 of 2)
>
AWGN
>
Real Time I/Q Baseband AWGN
3.
Press
Bandwidth
>
10
>
MHz
.
Configuring the RF Output
1.
Set the RF output frequency to 500 MHz.
2.
Set the output amplitude to
−
10 dBm.
3.
Press
RF On/Off
.
Generating the Waveform
Press
AWGN Off On
until On is highlighted.
This generates an AWGN waveform with the parameters defined in the previous procedure. During
waveform generation, the
AWGN
and
I/Q
annunciators activate. The waveform is now modulating the
RF carrier.