Basic Digital Operation
Using Waveform Markers
1. Clearing Marker Points from a Waveform Segment
When you set marker points they do not replace points that already exist, but are set
in addition
to
existing points. Because markers are cumulative, before you set points, view the segment (
page 102
)
and remove any unwanted points. With all markers cleared, the level of the event output signal is 0V.
Clearing All Marker Points
1.
In the Marker Utilities menu (
page 101
), press
Set Markers
.
2.
Highlight the desired waveform segment.
In an ARB format, there is only one file (
AUTOGEN_WAVEFORM
) and it is already highlighted.
3.
Highlight the desired marker number:
Press
Marker 1 2 3 4
.
The Set Marker display
These softkeys
change the range
of waveform
sample points
shown on the
marker display.
Each press of the
softkey changes
the sample range
by approximately
a factor of two
Marker
points on
first sample
point
First sample
point shown on
display
Select a segment
The display below shows the I and Q components of the waveform, and
the marker points set in a factory–supplied segment.
Summary of Contents for E8257D
Page 12: ...Contents xii ...
Page 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Page 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Page 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Page 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...