252
Chapter 12
Peripheral Devices
N5102A Digital Signal Interface Module
Figure 12-7
Clock Phase and Skew Adjustments
Connecting the Clock Source and the Device Under Test
As shown in
Figure 12- 3 on page 245
, there are numerous ways to provide a common frequency
reference to the system components (PSG, N5102A module, and the device under test).
Figure 12- 8
shows an example setup where the signal generator supplies the common frequency reference and the
N5102A module provides the clock to the device.
See the
N5102A Digital Signal Interface Module Installation Guide
for detailed information on
device interface connections.
CAUTION
The Device Interface connector on the interface module communicates using high speed
digital data. Use ESD precautions to eliminate potential damage when making
connections.
NOTE
You must disconnect the digital bus cable and the digital module while downloading
firmware to the signal generator.
Clock skew adjustment
90 degree phase adjustment
Clock
Phase adjusted
skew adjusted
clock
clock
Data
Phase and
Summary of Contents for E8257D