332
Chapter 6
Specifications
E5023A Specifications
Narrow Band TAA Measurement
PW (Pulse Width) Measurement
BL (Baseline) Measurement
Resolution Measurement
Absolute Measurement Accuracy:
For sinusoidal wave @Agilent E5038B input
±
6 % @ <600 MHz,
±
4 % (typical)
±
7 % @ 600 to 750 MHz
For any waveform @Agilent E5038B input
±
5 % (typical)
Table 6-34
Table 6-35
Parameters:
Amplitude of the basic frequency contents of the
read back signal (rms)
Data Pattern:
HF, LF, Isolated Pulse, Repetitive, NLTS-5th,
User defined pattern
Dynamic Range:
70 dB
Table 6-36
Parameters:
PW+, PW-, Average, PW asymmetry
Measurement Range:
1.5 nsec to 35 nsec
Data Pattern:
Isolated Pulse
Range of PW% (PWxx):
PW20 to PW80
Absolute Measurement Accuracy:
±
(10 % of r 50 psec)
±
(5 % + 50 psec) (typical)
Table 6-37
Parameters:
B, Baseline-, Average,
Baseline separation
Data Pattern:
Isolated Pulse
Measurement Accuracy:
±
5 % of TAAp-p (@ Input Range = Signal
Level)
Table 6-38
Parameters:
Resolution (%) [TAA
HF
/ TAA
LF
×
100 (%)]