hpe1437_frequency_setup
hpe1437_frequency_setup
sets all the zoomcenter frequency parameters. This
description also includes information on the following functions which set or query
frequency parameters individually:
hpe1437_frequency_cmplxdc
selects a complex baseband measurement
hpe1437_frequency_cmplxdc_get
gets the state of the baseband measurement mode
hpe1437_frequency_sync
prepares the module for a synchronous frequency change
hpe1437_frequency_sync_get
gets the state of the synchronous change mode
hpe1437_frequency_center
sets the center frequency
hpe1437_frequency_center_get
gets the current center frequency
VXI
plug&play
Syntax
#include “hpe1437.h”
ViStatus
hpe1437_frequency_setup
(ViSession
id
, ViInt16
cmplxDc
, ViInt16
sync
,
ViReal64
freq
);
ViStatus
hpe1437_frequency_cmplxdc
(ViSession
id
, ViInt16
cmplxDc
);
ViStatus
hpe1437_frequency_cmplxdc_get
(ViSession
id
, ViPInt16
cmplxDcPtr
);
ViStatus
hpe1437_frequency_sync
(ViSession
id
, ViInt16
sync
);
ViStatus
hpe1437_frequency_sync_get
(ViSession
id
, ViPInt16
syncPtr
);
ViStatus
hpe1437_frequency_center
(ViSession
id
, ViReal64
freq
);
ViStatus
hpe1437_frequency_center_get
(ViSession
id
, ViPReal64
freqPtr
);
Description
hpe1437_frequency_setup
sets the center frequency of a zoomed measurement. The
center of a frequency band of interest is converted to DC with this function. The
frequency transition is phase continuous unless the center frequency is set to zero in
which case the transition may be selected either to be phase continuous or phase reset.
This function may also be used to synchronously change frequency in multiple-module
systems.
Parameters
id
is the VXI instrument session pointer returned by the
hpe1437_init
function.
cmplxDc
selects either a phase continuous or phase reset transition when the
freq
= 0.
HPE1437_OFF
, combined with a frequency change to zero, causes phase to be reset to
zero.
HPE1437_ON
, combined with a frequency change to zero, does not reset the
phase, thereby generating a complex DC measurement at baseband. The state of this
parameter does not affect any transition where
freq
≠
0. Whether the real or complex
data is saved and ultimately sent to the output port is determined by the
hpe1437_data_type
function.
cmplxDcPtr
contains the value of the
cmplxDc
parameter.
E1437A User's Guide
VXI
plug&play
Programming Reference
4-33
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for E1437A
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 177: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 187: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 201: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 207: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...