hpe1437_filter_resp_get
Returns the module’s complex frequency response.
VXI
plug&play
Syntax
#include “hpe1437.h”
ViStatus
hpe1437_filter_resp_get
(ViSession
id
, ViReal64
resp
[ ], ViInt32
n
, ViReal64
fmin
, ViReal64
fmax
);
Description
This function uses the current filter and center frequency settings to return the complex
frequency response. The requested number of samples are equally spaced from the
requested minimum frequency to the requested maximum frequency.
Parameters
id
is the VXI instrument session pointer returned by the
hpe1437_init
function.
resp
returns the response in the format:
resp(re0, im0, re1, im1,..., re(n-1), im(n-1)
n
is the number of samples desired.
fmin
is the minimum frequency in Hertz.
fmax
is the maximum frequency in Hertz.
Effect on Active
Measurement
This command does not abort any measurement in progress.
See Also
hpe1437_init, hpe1437_filter_setup, hpe1437_frequency_setup
E1437A User's Guide
VXI
plug&play
Programming Reference
4-27
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for E1437A
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 177: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 187: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 201: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 207: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 209: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...