INPUT:RANGE:AUTO
command
Performs auto-ranging.
Command Syntax:
INPUT:RANGE:AUTO <sec>
sec::=<numeric>
numeric::=
≥
0 seconds
Description:
This command sets the range of a E1437 to the lowest value that will not cause an
ADC overload to occur. The algorithm will start at the lowest range and move up
until there is no ADC overload.
Parameter definitions:
is the time in seconds to take data at each range to insure that an overload is
detected. Setting this parameter to 0.0 will result in this time being set automatically
according to an algorithm that depends on block size and filter bandwidth.
NOTE
An autorange that is pending or in progress will be aborted if an INPUT:RANGE
or another INPUT:RANGE:AUTO command is received.
See Also
INPUT:SETUP
E1437A
INPUT:RANGE:AUTO
ASCII Overview and Commands
command
5-30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for E1437A
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 177: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 187: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 201: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 207: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 209: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...