Power Measurement
How to Measure Power
78
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
N O T E
If you see the screen shown in Figure 47, the zeroing operation has failed because
the Power Meter received light. The most common reason for zeroing to fail is if:
•
a source is connected to the Power Meter’s input connector,
•
the fiber connected to the Power Meter’s input connector is collecting light, or
•
the Power Meter receives ambient light because the input connector is
uncovered.
Press
Enter
and start again at step 1.
4
When the zeroing operation finishes, press the [Close] softkey to close the
menu.
Figure 46
Zeroing Screen