![Agilent Technologies Agilent 8163A User Manual Download Page 271](http://html.mh-extra.com/html/agilent-technologies/agilent-8163a/agilent-8163a_user-manual_2867997271.webp)
What do I need for proper cleaning?
Cleaning Instructions
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
271
Ideally, the light source on your microscope should be very flexible. This will
allow you to examine your device closely and from different angles.
A microscope helps you to estimate the type and degree of dirt on your device.
You can use a microscope to choose an appropriate cleaning method, and then to
examine the results. You can also use your microscope to judge whether your
optical device (such as a connector) is severely scratched and is, therefore,
causing inaccurate measurements.
Ultrasonic bath
Ultrasonic baths are also available from photography or laboratory suppliers or
specialist mail order companies.
An ultrasonic bath will gently remove fat and other stubborn dirt from your
optical devices. This helps increase the life span of the optical devices.
Only use isopropyl alcohol in your ultrasonic bath, as other solvents may damage.
Warm water and liquid soap
Only use water if you are sure that there is no other way of cleaning your optical
device without corrosion or damage. Do not use hot water, as this may cause
mechanical stress, which can damage your optical device.
Ensure that your liquid soap has no abrasive properties or perfume in it. You
should also avoid normal washing-up liquid, as it can cover your device in an
iridescent film after it has been air-dried.
Some lenses and mirrors also have a special coating, which may be sensitive to
mechanical stress, or to fat and liquids. For this reason we recommend you do not
touch them.
If you are not sure how sensitive your device is to cleaning, please contact the
manufacturer or your sales distributor.
Premoistened cleaning wipes
Use pre-moistened cleaning wipes as described in each individual cleaning
procedure. Cleaning wipes may be used in every instance where a moistened soft
tissue or cotton swab is applied.
Polymer film
Polymer film is available from laboratory suppliers or specialist mail order
companies.
Using polymer film is a gentle method of cleaning extremely sensitive devices,
such as reference reflectors and mirrors.