![Agilent Technologies Agilent 8163A User Manual Download Page 126](http://html.mh-extra.com/html/agilent-technologies/agilent-8163a/agilent-8163a_user-manual_2867997126.webp)
Tunable Lasers
How to Use Triggers
126
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
3
Move to <
Input Trigger Mode
>, by using the cursor key, and press E
NTER
.
You will see the screen in Figure 81.
4
Move to one of the following, by using the cursor key:
•
<Ignore
>, input triggers are ignored.
•
<
Next Step
>, an input trigger will cause the next step of a stepped sweep to be
performed.
•
<
Run Sweep
>, an input trigger will start a single sweep cycle.
5
Press
Enter.
You can generate input triggers in any of the following ways:
•
applying a trigger to the Input Trigger Connector on the rear panel of your
instrument,
•
setting <
Trigger Configuration
> to <
Loopback
> so that an output trigger
automatically generates an input trigger, or
•
using the
:TRIGger
GPIB command, see your mainframe’s Programming
Guide.
How to Use Output Triggering
You can configure your Tunable Laser module to output a trigger when the
instrument performs certain tasks.
To set your module’s Output Trigger Configuration:
1
See
“How to Set the Trigger Configuration” on page 62
for how to configure
the trigger connectors.
Figure 81
Input Trigger Mode