Performance Test Instructions
Performance Tests
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
251
24
Turn the Modify Knob counter clockwise.
Check the fields on the display being highlighted one after another.
25
Using the Modify Knob, move to the [
λ
] parameter of the power sensor (that
is, channel 2, right field).
26
Press the Modify Knob. The first digit left of the period should be highlighted.
27
Press [Cancel].
28
Press
Preset
, the green hardkey, to return to the Overview Screen.
Testing the Number Keys
N O T E
Number Keys are only available if you use the Agilent 8164A.
29
Using the cursor keys, move to the [P] parameter of the sensor channel, that is
channel 2. Press [Details].
30
Using the cursor keys, move to [
Cal
], and press
Enter
. The first digit left of the
period should be highlighted.
31
Using the number keys press the following sequence: 1, 2, 3, 4, 5, and 6.
123.456 dB should be displayed as the [
Cal
] value.
32
Press
Enter
twice to get access to the [
Cal
] field again
33
Using the number keys press the following sequence: 1, 2, ., 7, 8, and 9. 10.789
dB should be displayed as the [
Cal
] value.
34
Press [+/
−
] several times. The [
Cal
] field should toggle between 10.789 dB and
−
10.789 dB.
35
Press
Preset
, the green hardkey, to return to the Overview Screen.
Module Interaction Test
N O T E
The following functional test assumes the use of modules which are within
specification.
1
Setup the test equipment as shown in Figure 171:
Insert the fixed laser source module in slot 1 and the power sensor in slot 2 . If
you use the Agilent 8164A Lightwave Measurement System, insert the back-
loadable Tunable Laser module, if available, in slot 0.
2
Switch on the mainframe and wait until it has booted.
3
Press
Preset
, the green hardkey, to set all parameters to their default values.
4
Ensure that [
λ
], the wavelength of the power sensor module equals [
λ
], the
wavelength of the laser source in slot 1.
5
Ensure, the attenuation of the laser source is set to 0.00 dB.