TECHNICAL INFORMATION
If the
V
key is pressed the display shows:
Or:
M O D E :
v
A
U
T
O
M
A T I C
T S P F i
l a m e n t X
v
M A N
The message
MAN
at the end of the second row
can be different. It depends of the operating mode
selected:
In this operating mode only the sublimation current
value and the pressure threshold are set by means
of the control panel. The sublimation occurs when
the pressure threshold is reached (analog input on
pin 2 of the I/O connector, range 0 to 5 Vdc).
−
in mode AUTOMATIC the message become
AUT
If the
W
key is pressed the display shows the
previous message (mode MANUAL).
−
in mode REMOTE SET the message become
REM
If the
V
key is pressed the display shows:
−
in mode AUTOMAT/REMOTE the message
become
A/M.
M O D E :
v
R E M O T E S E T
NOTE
The controller escape automatically from the
operating mode selection before 30 second, if no
selection are made, and return to the normal
operation.
If the
MODE
key is pressed during the operating
mode selection the Sublimation Controller returns to
the normal operation and the operating mode
remains that previously selected before the
activation of the operating mode selection.
In this operating mode only the time and period of
sublimation are set by means of the control panel.
The current value of sublimation is set by means of
a signal incoming on the analog input on pin 2 of
the I/O connector, range 0 to 10 Vdc (0 Vdc = 0 A,
10 Vdc = 50 A).
If the
W
key is pressed the display shows the
previous message (mode AUTOMATIC).
If the
V
key is pressed the display shows:
Sublimation Parameters Setting
The mode by means of which is possible to modify
the sublimation parameters depends on the
selected operating mode (MANUAL, AUTOMATIC,
REMOTE SET and AUTOMAT/REMOTE). In the
following paragraph are described the different
setting for each operating mode.
M O D E :
v
A U T O M A T / R E M O T E
In this operating mode the sublimation starts when
the pressure threshold is reached (as in the
AUTOMATIC operating mode).
To display the sublimation parameters menu and to
switch from one page to another, press the
W
or
V
keys.
The current value of sublimation is set by means of
a signal incoming on the analog input on pin 2 of
the I/O connector, range 0 to 10 Vdc.
When the desired parameter is displayed, the value
can be changed as follows:
If the
W
key is pressed the display shows the
previous message (mode REMOTE SET).
1) press the
SET
key
2) press
V
key to increase the value or
W
key to
decrease the value
If the
V
key is pressed the display shows the first
message illustrated before (mode MANUAL).
3) press
SET
key to confirm the choice or
MODE
key to escape without changing the value. In any
case the controller come back to the normal
operating mode and the display shows one of
the following messages, depending on the
source type selected.
To select the desired operating mode press the
SET
key. The controller comes back to the normal
operating mode and the display shows one of the
following messages, depending on the source type
selected.
M i n i T i - b a l l
v
M A N
M i n i
T i - b a l l
v
M A N
21 87-900-077-01(D)
Summary of Contents for 929-0022
Page 2: ...Sublimation Controller ...
Page 3: ......
Page 4: ......