INSTRUCTION FOR USE
14 87-900-077-01(D)
Operating voltage selector
INSTALLATION
,
WARNING!
The Sublimation Controller is equipped with a 3-wire power cord
and plug (internationally approved) for user’s safety. Use this
power cord and plug in conjunction with a properly grounded
power socket to avoid electrical shock. Verify that the ground
connections are properly connected.
High voltages developed inside the controller can cause severe
injury or death. Before servicing the unit, disconnect the input
power cable.
CAUTION
The controller can be used as a bench unit or a rack module, but
it must be positioned so that free air can flow through the holes of
the top and side cover.
If the Sublimation Controller is used as a rack module, it MUST
be inserted in a 4 unit height rack adapter chassis to avoid
collapse of the controller fall inside the rack. The front panel of
the Sublimation Controller is not designed to support the
controller weight.
Do not install or use the Sublimation Controller in an environment
exposed to atmospheric agents (rain ,snow, ice), dust, corrosive
gases or in explosive environments or those with a high fire risk.
NOTE
If the Sublimation Controller is used as a bench unit, it must have
the four side legs. If the Sublimation Controller is installed on a
rack, remove the four side legs and position it with at least 30
mm (1,2 inches) of clearance on top and bottom.
During operation, the following environmental conditions must be
respected:
−
temperature: from +5 °C to +45 °C
−
relative humidity: 0 - 90% (non-condensing)
To connect the Sublimation Controller to the pump use the
specific cable supplied as an option. See the appendix
“Technical Information” for detailed information about the above
mentioned and other connection and about the optional card
installation.
USE
This paragraph describes the fundamental operating procedures.
Detailed information and operating procedures that involve
optional connections or options are supplied in the paragraph
“USE” of the appendix “Technical Information”.
Make all electrical connections of the controller and the pump
prior to operating the controller. Read carefully the pump
instruction manual prior to operating the system.
,
WARNING!
To avoid injury to personnel and damage to the equipment, if the
pump is laying on a table make sure it is secure.
Never operate the pump if the pump inlet is not connected to the
system or blanked off.
Sublimation Controller Front Panel Controls and
Indicators
The Sublimation Controller front panel is shown in the following
figure.
The control and indicator meanings and functions are detailed in
the following table.
Sublimation Controller front panel controls