INSTRUCTION FOR USE
15 87-900-077-01(D)
Ref. Control/Indicator
Name
Control/Indicator Description
1
LCD back lighted alphanumeric display dot matrix, 16 characters, 2 rows.
2 TSP
FILAMENT
1 , 2, 3
Green LEDs that show:
•
if ON, the relevant TSP pump filament selected;
•
if blinking, the relevant TSP pump filament interrupted.
3
SEL
Key for manual selection of one TSP pump filaments.
4
ON/OFF
Key for controller operating mode selection:
•
OFF when “Parameters Setting” mode is activated;
•
ON when “Sublimation” is activated.
5
SUBLIMATION
Green LED that show:
•
if ON, that controller status is in waiting time between two sublimation or during current leading
edge;
•
if blinking, the sublimation is in phase.
6
MINI TIBALL
Green LED ON when Mini Ti-Ball operating mode is selected.
7
SET
Key that, depending on condition:
•
confirm the start of sublimation;
•
memorize the selection made;
•
reset a fail status.
8
MODE
Key that, depending on condition:
•
select the operating mode (MANUAL, AUTOMATIC, AUTOMAT/REMOTE and REMOTE SET);
•
escape from the parameters setting pages without changing the parameters.
9
[
V
/
W
]
Keys used to modify the values in the parameters setting pages.
Sublimation Controller Rear Panel
The following figure shows the controls and the connections of
the Sublimation Controller rear panel.
OPERATING PROCEDURES
Operating Modes
The Sublimation Controller can be operated in the MANUAL,
AUTOMATIC, REMOTE SET and AUTOMAT/REMOTE mode.
In MANUAL and AUTOMATIC modes all the controls are given
on front panel.
In AUTOMAT/REMOTE and REMOTE SET, the controls are
given via signals coming through the I/O cable.
NOTE
After a power failure the Sublimation Controller will return to the
previously selected operating mode. If “AUTOSTART” parameter
was set on “YES” and before the power failure the unit was in
sublimation mode, the sublimation starts on again automatically.
The unit is designed to operate in two different modes:
•
with manual selection of the parameters (
MANUAL
and
REMOTE SET
modes)
•
with automatic selection of the parameters (
AUTOMATIC
and
AUTOMAT/REMOTE
modes)
The Sublimation Controller can be used to control the TSP or
Mini Ti-Ball pumps type.
In MANUAL and REMOTE SET operating modes the sublimation
parameters are setting by the user by means of the front panel
controls (MANUAL mode) and/or by signal send on the I/O
connector (REMOTE mode).
If the Sublimation Controller is used in conjunction with the TSP
pump the TSP source type must be set. If the Sublimation
Controller is used in conjunction with the Mini Ti-Ball pump the
Mini Ti-Ball source type must be set.
Sublimation Controller rear panel
NOTE
The unit is factory set as follows:
−
MANUAL operating mode
−
TSP source type
−
sublimation current 30 A
−
sublimation time 1 minute
−
sublimation period 3 minutes
−
pressure threshold 1x10
-7
mbar
−
“AUTOSTART” (automatic restart before power failure) on
“YES”
1. Connector to plug the optional cable for TSP or Mini Ti-Ball
connection
2. RS 232-422-485 connector (optional)
3. Power cord socket, input voltage selector, protection fuses
4. Power switch: ON (I)/OFF(O)
5. I/O signals connector