TECHNICAL INFORMATION
19 87-900-077-01(D)
USE
There are three different type of settings that can be
selected;
1) configuration of the Sublimation Controller
2) operating mode selection
3) parameters setting.
NOTE
The unit is factory set for:
−
MANUAL mode operation
−
source TSP
−
sublimation current 30 A
−
sublimation time 1 minute
−
sublimation period 3 minutes
−
pressure threshold 1x10
-7
mbar
−
“AUTOSTART” set YES (automatic restart after
shut down)
NOTE
The message sequences illustrated below is
referred at the first operation of the Sublimation
Controller with the default settings.
Configuration Selection
The Sublimation Controller configuration permits
selection of the follows aspects:
•
type of source that can be used by Sublimation
Controller
•
enables/disables of automatic restart of
sublimation after a shut down
•
the automatic/manual selection of the TSP
filament for the sublimation
To enable the configuration mode that permits the
selection of the source type press the
MODE
and
SET
keys, on the front panel, and simultaneously,
on rear panel, switch on the power. The display
shows:
S
O
U
R
C
E T
Y P
E
:
T S P
Press
V
or
▼
to change the source type from
TSP to Mini Ti-Ball, and the display shows:
S O U R C E T Y P E
:
M i n i T i
- b a l l
Press
SET
to confirm the choice. For a few seconds
the display shows, depending from the selection,
one of the following message:
S O U R C E S E L E C T E D
T S P
Or:
S O U R C E
S E L E C T E D
M i
n
i
T i - b a l l
Then the controller passes to the normal start up
operations and the display shows:
A U T O
T
E
S
T
Then, if a serial module is mounted on the
controller, the display shows:
S E R I A L
T E S T
At the end of this check, if no problems on the serial
port are detected, the display shows:
S E R I A L
O K
Before the serial port check, if the Computer
Interface Card is mounted or before the message
“AUTOTEST” the software check the presence of
the zero-crossing signal and, in case of fault
(frequency absent or different from 50 or 60 Hz),
the display shows:
P O W E R
A C F A I L
and the controller must be shut down.
If the check has a positive result, the display shows,
for a few seconds:
A U T O
T
E
S
T
O K
Then the display shows one of the following
message, depends of the source type previously
selected:
M i
n
i
T
i
- b a l l
v