Performance Tests
Performance Test Instructions
66
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
N O T E
To ensure traceability, use the 81610CC Reference Cable for
calibration measurements.
Do not
use the 81610CC Reference Cable for measurements on a
Device Under Test. Instead, use a measurement patchcord.
C A U T I O N
It is important to maintain the quality of the straight connector end of
the 81610CC Reference Cable.
Never
add another connector to the straight end of the 81610CC
Reference Cable, since a physical connection is made.
When the straight end of the 81610CC Reference Cable is connected to
the 81634A Power Sensor module, no physical connection is made, so
there should be no degradation of connector quality.
10
Enter the power [
P
] reading on the Power meter as [
FP Delta
] on
the Return Loss meter.
Make sure you enter the sign of this value correctly.
11
At the optical attenuator:
a
Set ‘Attenuation’ to 60.0 dB
b
Set ‘Wavelength’ to the wavelength of the ASE source
c
Enable the attenuator.
12
At the Return Loss meter, press [TermCal] to calibrate the Return
Loss module at termination condition.
13
At the optical attenuator, set the ‘Attenuation’
to 0.0 dB
The measured Return Loss should now be identical to the [
RLref
]
value entered at step 7.
14
Note this return loss value in the Test Record.
15
At the optical attenuator, increase the ‘Attenuation’ setting until the
Return Loss reading matches the next value in the Test Record.
Note this ‘Attenuation’ setting in the Test Record as ‘Actual
Attenuation’.
16
Repeat steps 13 and 14 for each value in the test record.
17
For each Return Loss reading, calculate the result by:
a
Calcuating the Effective Attenuation by multiplying the Actual
Attenuation by 2.
b
Subtracting the RLref value from the Return Loss value, then
subtracting the Effective Attenuation value.
An example follows:
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......