Getting Started with Return Loss
Making a Return Loss Measurement
30
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Figure 22
Measuring the Termination Parameter - External Source
or Figure 23 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 23
Measuring the Termination Parameter - Internal Source
2
Terminate the measurement patchcord so that there are no
reflections coming from the end.
T I P
You can do this by wrapping the fiber five times around the shaft of
a screwdriver
(or some similar object with a diameter of between
5mm and 7mm).
3
Move to <
Terminated calibration
> and press
Enter
. The instrument
measures the power reflected by the component, and sets the
[
Para
] values used by the Return Loss monitor’s power sensor and
monitor diode.
8161x
8163A Lightwave Multimeter
Termination
measurement patchcord