Getting Started with Return Loss
Making a Return Loss Measurement
26
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Calibration using the Agilent 81610CC
Reference Cable
Use the Agilent 816100CC Refernce Cable to make a Reflectance
Calibration, and the measurement patchcord to make a Termination
Calibration.
To measure insertion loss and the front panel delta of the system (see
“Calculating the Front Panel Delta” on page 41
), you must also
measure the insertion loss of the Reference Cable and measurement
patchcord.
Reflectance Calibration
Use this procedure to calibrate the Return Loss module against a
component of known return loss.
1
Attach the source to the Return Loss module’s input as shown in
Figure 16. Attach the high return loss connector of the reference
cable to the output.
Figure 16
Measuring the Return Loss of the Reference Cable - External source
or Figure 17 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 17
Measuring the Return Loss of the Reference Cable - Internal Source
Attach the high return loss connector of the reference cable to the
output.
T I P
For best results and higher repeatability, fix the cable.
2
Move to [
FPDelta
], press [Edit]
,
set the value to 0.000, then press
[OK].
Agilent 81610CC Reference Cable
8161xA
8163A Lightwave Multimeter
Laser
Source
Agilent 81113PC
8161x
8163A Lightwave Multimeter
Agilent 81610CC Reference Cable
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......