Getting Started with Return Loss
A Background to Return Loss Measurement
38
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
Measuring the Power Transmitted Through
the Measurement Patchcord
Connect the measurement patch cord (the cable you will use to
connect to the Device Under Test, DUT) directly to a Power Meter and
measure the transmitted power,
E
Meas
. You can use
E
Meas
to calculate
the front panel delta, see
“Calculating the Front Panel Delta” on
and to calculate the insertion loss, see
Insertion Loss of the DUT” on page 43
Figure 32
Measuring the Power transmitted through the Measurement Patchcord
Measuring the Reflections from the DUT
Now detach the measurement patchcord from the Power Meter and
attach it to the DUT. The DUT should be terminated.
Figure 33
Measuring the Reflections from the Device Under Test
The instrument measures the power reflected from the DUT. This
power is called
P
DUT
.
Power Meter
E
Meas
M
ref
P
ref
Agilent 81610A
Reflection Reference
measurement patchcord
M
dut
P
dut
DUT
Agilent 81610A
Reflection-free
Termination
measurement
patchcord
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......