Getting Started with Return Loss
A Background to Return Loss Measurement
36
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
A Background to Return Loss
Measurement
Measuring the Reflected Power from a
Component with Known Reflection Factor
First, attach a component with a known reflection factor in place of
the DUT, and measure the power reflected. This is component is called
the reflection reference.
Figure 29
Measuring the Power from a Component with a Known Reflection Factor
This measured power from the reflection reference is called
P
Ref
.
The following two components are recommended for use as a
reflection reference:
• the Agilent 81610CC Reference Cable, or
• the Agilent 81000BR Back Reflector.
N O T E
Note you can only measure the Front Panel Delta if you use a
Reflection Reference Cable.
The reflection factor for the component is called
R
R
. Normally the
return loss for the component (
RL
R
) is specified, but these values are
related:
Reflection
Reference
R
ref
M
ref
P
ref
Agilent 81610A
Reflectivity
P
in
P
r
5/
5
10
5
log
5
–
10
3
U
3
LQ
-------
log
–
=
=
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......