Performance Test Instructions
Performance Tests
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
69
6
At the Power meter:
a
Set the averaging time [Av Time] to 1s
b
Set the wavelength [l] to the wavelength of the RTL source.
c
Set [Pwr unit] to <dB>.
7
At the return loss meter:
a
Set the averaging time [Av Time] to 1s.
b
Enter the Return Loss Reference value RLref of the 81610CC
reference cable for this wavelength. Note this value in the Test
Record as RLref.
c
Press [RefCal] to calibrate the Return Loss module at reference
condition.
8
At the powermeter press [Disp®Ref]. The Power meter should now
read 0.0dB.
9
Setup the equipment as shown in figure D-4. Use an special
pigtailed optical attenuator w/ an angled input and straight output
connector and connect the angled connector to the output of the
81611-14A return loss module. The output fiber of the optical
attenuator is connected to the input of the powermeter. All optical
cables used in this setup should be fixed.
Figure 40
Relative Uncertainty of Return Loss - Performance Test Setup
10
Enter the power [P] reading on the Power meter as [FPDelta] on the
Return Loss meter. Make sure you enter the sign of this value
correctly.
11
At the optical attenuator:
a
Set 'Attenuation' to 60.0dB.
Mainframe 8163A w/ 8161xA RTL
and 81634A Power Sensor
8163A
8161xA
in
out
81634A
fiber
pigtail
Optical Attenuator 8156A
8156A #special
fiber
pigtail
pigtailed version
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......