Performance Test Instructions
Performance Tests
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
65
6
At the Power meter:
a
Set the averaging time [
Av Time
] to 1 s
b
Set the wavelength [
λ]
to the wavelength of the ASE source.
c
Set [Pwr unit] to <
dB
>.
7
At the Return Loss meter:
a
Set the averaging time [
Av Time
] to 1 s
b
Set the wavelength [
λ]
to the wavelength of the ASE source.
c
Enter the Return Loss Reference value [
RLref
] of the 81610CC
reference cable for this wavelength. Note this value in the Test
Record as RLref
d
Press [RefCal] to calibrate the Return Loss module at reference
condition.
8
At the Power meter, press [Disp
→
Ref].
The Power meter should now read 0.0 dB.
9
Setup the equipment as shown in Figure 38. Use an optical
attenuator with angled connectors, and connect the angled
connector of the 81610CC Reference Cable to the output of the
attenuator.
The optical cable from the laser source, and the optical cables to
and from the 8156A Attenuator, must be fixed to the table to ensure
minimal cable movement during the tests.
Figure 38
Relative Uncertainty of Return Loss - Performance Test Setup
83438A ASE Source
83438A
Mainframe 8163A w/ 8161xA RTL
and 81634A Power Sensor
8163A
8161xA
in
out
81634A
81113PC
81610CC
Optical Attenuator 8156A
8156A #201
81113SC
angled conectors
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......