59
Programming Reference
SCPI Instrument Command List
Command
:PULS:DEL[1]:UNIT
Long
[:SOURce]:PULSe:DELay[1]:UNIT
Form
Set & Query
Parameter
S | SEC | PCT | DEG | RAD
*RST value
S
Description
Use this command to set/read the default units for the pulse delay
parameter. The default unit of a parameter is the unit used when the
parameter is programmed to a value without a unit suffix.
Example
To set the pulse delay to 50% of period:
Command
:PULS:DOUB[1]
Long
[:SOURce]:PULSe:DOUBle[1][:STATe]
Form
Set & Query
Parameter
OFF | ON
*RST value
OFF
Description
Use this command to switch double-pulse mode on or off. In double-
pulse mode two pulses are generated per pulse period. The delay
between the leading edges of the first and second pulse can be adjusted.
:PULS:DEL:UNIT PCT
:PULS:DEL 50
Set OUTPUT delay unit to %
Set OUTPUT delay to 50% of period