Programming Reference
SCPI Instrument Command List
•
Specified Voltage limits
•
Actual OUTPUT Impedance setting
:OUTPut:IMPedance
•
Actual Expected Load impedance setting:
:OUTPUT:IMPedance:EXTernal
Example
To program the low level current of the output signal:
Command
:CURR[1]:LIM
Long
[:SOURce]:CURRent[1]:LIMit[:HIGH]
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
+10.0 mA
Description
Use this command to set/read the high level current limit. If you switch
on current limiting, the high level current cannot be set above the
programmed limit.
The current is
not
limited by the OUTPUT hardware, this is a software
limit.
Example
To set the high level current limit for the output signal:
:HOLD CURR
:CURR:LOW 50 MA
Enable CURRENT subsystem
Set OUTPUT low level to 50 mA
:HOLD CURR
:CURR:LIM 50 MA
:CURR:LIM:STAT ON
Enable CURRENT subsystem
Set OUTPUT high level current limit to 50 mA
Switch on OUTPUT limits